Proven performance for cost-effective sterilization using ASP's original gas plasma technology STERRAD® 100S continues to be the system of choice for facilities looking for a proven, reliable, and cost-effective low-temperature sterilization solution:
Consult the STERRAD® Sterility Guide, www.sterradsterilityguide.com, refer to the STERRAD® Sterilization System User's Guide or contact the medical device manufacturer for detailed processing information.
ASP offers the STERRAD® 100S System, a proven cost-effective sterilization solution.
Safely and effectively sterilizes a wide range of instruments, including flexible endoscopes with lumens made from polyethylene or Teflon® and having an inside diameter of ≥ 1 mm and a length of ≤ 500 mm or an inside diameter of ≥ 1 mm and a length of > 500 mm but ≤ 2000 mm.
Technical Specifications | ||||
Total Cycle Time | Short Cycle 54 minutes Long cycle 72 minutes |
|||
Cycle Temperature | Less than 55°C | |||
Dimensions and Weight | ||||
Height | 1664 mm | |||
Width | 7625 mm | |||
Depth | 1046 mm | |||
Weight | 350 kg | |||
Sterilization Chamber | ||||
Configuration | Circular | |||
Chamber volume | 173 L * | |||
Shelf Dimensions | Upper shelf:700 mm x 425 mm Lower shelf: 760 mm x 325 mm |
|||
Installation and Electrical Requirements | ||||
Placement | Freestanding with built-in wheels for mobility | |||
Connections | 380-415 VAC, 50 Hz , 10 A, 3-phase circuit with separate neutral and ground conductors. CBA phase rotation | |||
Service Requirements | In operation, the STERRAD® 100S System should not be placed closer than 5 cm to a wall at the rear and/or sides of the system. The sterilizer should be installed in a space of sufficient size to permit access to all 4 sides of the system when it is rolled from the rear and/or side walls a distance of 3 m. Service access requires a minimum clearance of 1 m on all sides of the system. The power receptacle should be positioned 30.5 cm to 61 cm above the floor. | |||
Operational Environment | ||||
Temperature | 10°C to 40°C | |||
Humidity | 30-75% relative humidity (non-condensing) |
*Please contact your ASP representative to determine maximum capacity based on your facility's particular sterilization load configuration(s).