With the STERRAD® NX® System, terminally sterilized instruments are readily available to help you keep your OR ahead of schedule while providing the highest level of sterility assurance for your patients.
Thank you for choosing the STERRAD® NX® System for your facility! In the near future, a highly trained ASP field service engineer will be on site to install your STERRAD® NX® System. The following are some pre-installation considerations that will assist in making your installation experience as seamless as possible.
STERRAD® NX® System Location | YES | |
Using STERRAD® NX® System cart or a countertop that meets specification | ||
24-inch clearance above unit | ||
24-inch clearance to right of unit | ||
System is at least 1 inch from the wall at the rear | ||
Easy access to and removal of all four sides | ||
Power receptacle is 12 to 24 inches above the floor | ||
Electrical Assessment | ||
Single phase dedicated circuit with dedicated hot and neutral wires, and a NEMA 5-20 receptacle with a 20 amp circuit breaker. Wire gauge sufficient to maintain 108-132 VAC with 16 amps continuous and 40 amps momentary current. Please see additional requirements on previous page under the heading CONDUCT AN ELECTRICAL ASSESSMENT. | ||
SS1 Removed | ||
System disconnected and removed from area | ||
Water filtration devices and plumbing removed | ||
Alternate Sterilization Location | ||
Plans confirmed for alternative sterilization location for 48 hours |