The Fast-Acting, Reliable Glutaraldehyde Solution

For disinfection and sterilization of instruments and endoscopes requiring fast turnaround, CIDEXPLUS® 28 Day Solution gives you the flexibility to run your facility efficiently.

  • 3.4% alkaline glutaraldehyde solution provides quick high-level disinfection 20 minutes
  • Destroys all vegetative forms of bacteria and 100% of Mycobacterium tuberculosis in
    20 minutes at 25°C; effective in the presence of 2% organic soil
  • Compatible with a wide range of medical instruments

Using CIDEXPLUS® 28 Day Solution Test Strips

To verify the maximum efficacy of your CIDEXPLUS® 28 Day Solution, ASP recommends testing for the minimum effective concentration (MEC) prior to each use. ASP offers test strips designed specifically for use with CIDEXPLUS® 28 Day Solution. Easy to use and read, CIDEXPLUS® 28 Day Solution Test Strips are the most reliable way to verify that CIDEXPLUS® 28 Day Solution will destroy microorganisms, including the highly resistant Mycobacterium tuberculosis .

CIDEXPlus Test Strip Color Chart
Reorder# Product Description Case Contents
2924 CIDEXPLUS® Test Trips Test Strips 2 Bottles - 60 Strings per bottle

Quality Assurance Testing Frequency

It is recommended that the testing of positive and negative controls be performed on each newly opened test strip bottle of CIDEXPLUS® 28 Day Solution Test Strips. After this initial testing, it is recommended that testing of freshly prepared positive and negative controls be performed on a regular basis as established by your own quality control procedures and program. This testing program will serve to minimize errors between different users, use of outdated materials or product that has been improperly stored or handled.

For optimal results, please carefully follow Instructions for Use when using test strips:

  1. Completely submerge indicated pad of the strip into solution. Hold for 1 second then remove. Do not shake strip.
  2. Remove excess solution by standing the strip upright on a paper towel.
  3. Read results at 90 seconds. Do not read past 90 seconds. Pad will be completely red to indicate effective solution. If any yellow remains on the pad apart from the top line, the solution is ineffective and must be discarded.
  4. Tightly recap test strip bottle after each use.
  5. Discard any unused test strips 90 days after bottle is originally opened.

For more information, please see instructions for test strip use.

Download a printable version of the step-by-step instructionsfor CIDEXPLUS® 28 Day Solution.