Cycles |
Standard: sterilizes most general surgical instruments
Flex: sterilizes up to 2 single-channel flexible endoscopes
Express: sterilizes da Vinci® 3-D endoscopes and other delicate instruments without lumens
Duo: sterilizes cameras and up to 2 single-channel flexible endoscopes |
Time |
Standard: 47 minutes Flex: 42 minutes Express: 24 minutes Duo: 60 minutes |
Lumen claims |
Standard: diameter: ≥0.7 mm; length: ≤500 mm
Flex: diameter: ≥1 mm; length: ≤850 mm Express: nonlumen sterilization only
Duo: diameter: ≥1 mm; length: ≤875 mm |
Cycle temperature |
47°C-56°C |
Sterilant |
Hydrogen peroxide |
Sterilant delivery |
Delivered in closed system in cassettes with automatic detection of expiration date:
Standard/Express/Flex: 10.8 mL per cycle (double kill) (2 ampules, each ampule is 5.4 mL)
Duo: 3.1 mL per cycle |
Used cassette disposal |
Automatic and touchless ejection into cassette disposal container |
Peroxide residual breakdown |
Gas plasma technology breaks down H2O2 to safe elements of water and oxygen |
Configurations |
Single and double doors |
System dimensions (maximum) |
Height: 1800 mm Width: 775 mm
Depth: single door: 1055 mm; double door: 1095 mm |
Chamber total volume |
152 L |
Chamber dimensions |
Height: 410 mm Width: 510 mm Depth: 735 mm |
Chamber shape |
Rectangular |
Shelf info |
Two-tiered shelf: Width: 444 mm Depth: 643 mm |
Mobility |
4 casters (2 locking) |
System weight |
1 door: 382 kg 2 door: 408 kg |
User interface |
Touchscreen technology: projected capacitive touch Resolution: 800 x 600 pixels |
Supported USB devices |
• Barcode reader • External drives: USB that allows data upload and download
• External printer connection (PCL3 compatible or equivalent) |
Standards/Compliance |
ISO 14937 |
Sterilization process |
Terminal-sterilization, double-kill cycle to provide a Sterility Assurance Level (SAL) of 10-6; 2 injections and identical plasma phases |
Delivered sterilant concentration |
58%-59.6% |
Sterilization cycle monitoring |
Critical system parameters monitored with on-board sensors, biological indicators, and chemical indicators; IMS (independent monitoring system) available |
H2O2 concentration continuous monitoring |
Monitoring using UV sensor within the chamber |
Heat generation |
Idle: 4574 BTU/hour Maximum: 6187 BTU/hour |
Electrical power specifications |
• US/Canada: 208 VAC, 60 Hz, NEMA L21-30 5-wire grounding twist-lock outlet attached to a dedicated 30-amp, 3-phase wye configuration circuit with separate neutral and ground conductors
• International: 380-415 VAC, 50/60 Hz, 5-wire grounding outlet attached to a dedicated 30-amp, 3-phase wye configuration circuit with separate neutral and ground conductors
• Japan: 200 VAC, 50/60 Hz: 4-wire grounding outlet attached to a dedicated 30-amp, 3-phase delta configuration with separate ground conductors |
Installation space requirements |
In operation, the STERRAD® 100NX System with ALLClear™ Technology should not be placed closer than 50.8 mm from the rear wall. The power receptacle should be positioned 30.5 to 61 cm above the floor. For recessed systems, a clearance of 25 mm from the top of the system and 45 mm on each side of the system is required. |
Service space requirements |
Service access requires a minimum clearance of 610 mm above the top and approximately 991 mm in on all sides of the system (can be less if the system can be moved for servicing). |
Temperature |
18°C-35°C |
Humidity |
10%-85% RH (noncondensing) |
System performance data and reports |
Cycle history, full 1-second data files, and reports available via ASP ACCESS™ Technology |
Network connectivity |
Communication protocol for Instrument Tracking Systems (ITS) available via ASP ACCESS™ Technology |
Data recording |
• Electronic data storage up to 200 cycles • Internal printer for manual recordkeeping
• Full electronic cycle data and reports via ASP ACCESS™ Technology |
Code |
Product Description |
10104-005 |
STERRAD® 100NX Sterilization System with ALLClear™ Technology, single-door unit, STANDARD & FLEX cycles, includes installation |
10104-006 |
STERRAD® 100NX Sterilization System with ALLClear™ Technology, double-door unit, STANDARD & FLEX cycles, includes installation |
10104-007 |
STERRAD® 100NX Sterilization System with ALLClear™ Technology, single-door unit, STANDARD, FLEX and DUO cycles, includes installation |
10104-008 |
STERRAD® 100NX Sterilization System with ALLClear™ Technology, double-door unit, STANDARD, FLEX and DUO cycles, includes installation |
10144 |
STERRAD® 100NX System Cassette (2 cassettes/case) |
20227 |
Cassette Disposal Box (10 boxes/case) |
10135 |
STERRAD® 100NX System EXPRESS Cycle Kit |
10137 |
STERRAD® 100NX System DUO Cycle Kit |
10305 |
Thermal Printer Paper (12 rolls/case) |
113617-01 |
STERRAD® Systems Bar Code Scanner Kit |