The STERRAD® 100NX™ Sterilizer is designed for sterilization of both metal and non-metal medical devices at low temperatures. Because the cycle operates within a dry environment and at low temperatures, it is ideal for instruments sensitive to heat and moisture. The STERRAD® 100NX™ Sterilizer can also sterilize instruments which have diffusion-restricted spaces, such as the hinged portion of forceps and scissors.
Medical devices with the following materials and dimensions can only be processed in the STERRAD® 100NX™ Sterilizer Standard Cycle:
Single channel stainless steel lumens with:
An inside diameter of 0.7 mm or larger and a length of 500 mm or shorter
Polyethylene and Teflon lumen tubing (excluding flexible endoscopes) with an inside diameter of 1 mm or larger and length of 1000 mm or shorter. [These items must be sterilized without any additional load.]
Up to twenty (20) pieces of tubing may be sterilized at a time.
Many flexible endoscopes, with the following materials and dimensions can only be processed in the STERRAD® 100NX™ Sterilizer Flex Cycle:
Single channel polyethylene and Teflon flexible endoscopes with a diameter of 1 mm or larger and length of 850 mm or shorter
One or two flexible endoscopes may be sterilized at one time. These items must be sterilized without any additional load.
For further information on claims please see Chapter 8 of the STERRAD® 100NX™ System User's Guide.