Designed with the Next Generation technology, the STERRAD® NX™ System is ASP's compact size sterilizer ideal for use anywhere in the hospital:
Consult the STERRAD® Sterility Guide, www.sterradsterilityguide.com, refer to the STERRAD® Sterilization System User's Guide or contact medical device manufacturer for the detailed processing information
Average Cycle Time | 28/38 minutes Standard/Advanced cycles |
|||
Temperature | 18°C-35°C | |||
Height Width Depth |
838.2mm 558.8 mm 812.8mm |
|||
Weight | 125 kg | |||
Sterilization Chamber Configuration | Rectangular | |||
Chamber Volume | 51.3 L * | |||
Dimension | 157 mm x 320 mm x 600mm (HxWxD) | |||
Two-Tiered Shelf | 313 mm x 597 mm (WxD) | |||
Installation & Electrical Requirements | ||||
Door Width (Radius)/Angle | 558.8 mm/100° | |||
Cart with Locking Wheels | ||||
Cart Dimension (H/D/W) | 560 / 735 / 610 mm | |||
Cart weight | 38 kg | |||
Chamber | ||||
Placement | Countertop or cart-mounted for mobility | |||
Connections | 180-264 VAC, 47-63 Hz, 10 A | |||
Service Requirements | In operation, the STERRAD® NX™ System should not be placed closer than 25 mm to a wall at the rear. The sterilizer should be installed in a space of sufficient size to permit access to all 4 sides of the system. Service access requires a minimum clearance of 61 cm above the top and 61 cm to the right of the system. The power receptacle should be positioned 30.5 cm to 61 cm above the floor. | |||
Operational Environment Temperature | 18°C to 35°C | |||
Humidity | 10% to 85% up to 30°C, linearly decreasing from 85% at 30°C to 70% at 40°C | |||
Cart with Locking Wheels | ||||
Height Width Depth |
560 mm 610 mm 735 mm |
|||
Weight | 38 kg | |||
Independent Monitoring System (IMS) | Enables remote, independent monitoring of all processes and is available as an option | |||
Barcode Reader | Available as an option |
*Please contact your ASP representative to determine maximum capacity based on your facility's particular sterilization load configuration(s).