Designed with the Next Generation technology, the STERRAD® 100NX™ System is the most advanced hydrogen peroxide gas plasma sterilizer from ASP:

  • Large rectangular chamber and several cycle options for the most productivity 47-minute Standard Cycle for most surgical instruments.
  • 42-minute Flex Cycle for single-channel flexible endoscopes
  • 24-minute Express Cycle for fast turnaround of da Vinci® 3-D endoscopes*, rigid telescopes, rechargeable batteries, and many other instruments.
  • Advanced features for ease of use and quality control including large touch-screen display, innovative foot pedal to facilitate the loading process, hydrogen peroxide monitor to ensure that sufficient hydrogen peroxide is available to assist in achieving a sterility assurance level (SAL) of 10-6, network connectivity for remote cycle monitoring and many others.
  • Available as a pass-through sterilizer for the Central Sterilization Departments with a need for separation of clean and sterile areas.

Please refer to the STERRAD® 100NX™ User's guide for more detailed processing information.

* da Vinci® 3-D endoscopes are validated for use in the STERRAD® 100NX™ System EXPRESS cycle and are not validated for use in the STERRAD® 100NX™ System STANDARD and FLEX cycles.

Consult the STERRAD® Sterility Guide, www.sterradsterilityguide.com, refer to the STERRAD® Sterilization System User's Guide or contact medical device manufacturer for the detailed processing information.

STERRAD® 100NX™ System Specifications

Technical Specifications
Total Cycle Time Standard Cycle: 47 minutes Flex Cycle: 42 minutes Express cycle: 24 minutes
Cycle Temperature Less than 55°C
Dimensions and Weight
Height 1791 mm
Width 775 mm
Depth 1016 mm
Weight 413 kg single-door
436 kg double-door
Installation and Electrical Requirements
Connections 3-phase, 380-415 VAC, 16 A, 50 Hz
Service Requirements In operation, the STERRAD® 100NX™ System should not be placed closer than 5.1 cm from the rear wall. The sterilizer should be installed in a space of sufficient size to permit access to all 4 sides of the system. Service access requires a minimum clearance of 61 cm above the top and 20 cm on each side of the system. The power receptacle should be positioned 30.5 cm to 61 cm above the floor.
Operational Environment
Temperature 18°C to 35°C
Humidity 10% to 85% up to 30°C, linearly decreasing from 85% at 30°C to 70% at 40°C (non-condensing)
Independent Monitoring System (IMS) Enables remote, independent monitoring of all processes and is available as an option
Barcode Reader Available as an option

* Please contact your ASP representative to determine maximum chamber capacity based on your facility's particular sterilization load configuration(s).