Designed with the Next Generation technology, the STERRAD® 100NX™ System is the most advanced hydrogen peroxide gas plasma sterilizer from ASP:
Please refer to the STERRAD® 100NX™ User's guide for more detailed processing information.
* da Vinci® 3-D endoscopes are validated for use in the STERRAD® 100NX™ System EXPRESS cycle and are not validated for use in the STERRAD® 100NX™ System STANDARD and FLEX cycles.
Consult the STERRAD® Sterility Guide, www.sterradsterilityguide.com, refer to the STERRAD® Sterilization System User's Guide or contact medical device manufacturer for the detailed processing information.
Technical Specifications | ||||
Total Cycle Time | Standard Cycle: 47 minutes Flex Cycle: 42 minutes Express cycle: 24 minutes | |||
Cycle Temperature | Less than 55°C | |||
Dimensions and Weight | ||||
Height | 1791 mm | |||
Width | 775 mm | |||
Depth | 1016 mm | |||
Weight | 413 kg single-door 436 kg double-door |
|||
Installation and Electrical Requirements | ||||
Connections | 3-phase, 380-415 VAC, 16 A, 50 Hz | |||
Service Requirements | In operation, the STERRAD® 100NX™ System should not be placed closer than 5.1 cm from the rear wall. The sterilizer should be installed in a space of sufficient size to permit access to all 4 sides of the system. Service access requires a minimum clearance of 61 cm above the top and 20 cm on each side of the system. The power receptacle should be positioned 30.5 cm to 61 cm above the floor. | |||
Operational Environment | ||||
Temperature | 18°C to 35°C | |||
Humidity | 10% to 85% up to 30°C, linearly decreasing from 85% at 30°C to 70% at 40°C (non-condensing) | |||
Independent Monitoring System (IMS) | Enables remote, independent monitoring of all processes and is available as an option | |||
Barcode Reader | Available as an option |
* Please contact your ASP representative to determine maximum chamber capacity based on your facility's particular sterilization load configuration(s).