Speed and versatility in a compact size
Designed with the Next Generation technology, the STERRAD® NX™ System is ASP's compact size sterilizer ideal for use anywhere in the hospital:
- Two tiered 30 liter chamber and two cycle options for increased productivity:
- 28-minute Standard cycle for most surgical instruments
- 38-minute Advanced cycle for single-channel flexible endoscopes
- Space-efficient STERRAD® NX™ System can be easily cart-mounted and moved where needed
- Provides terminally sterile, packaged instruments thereby eliminating the need to compromise with liquid biocides
- Features & Benefits
- Specifications
- Downloads
STERRAD® NX™ System Specifications
Technical Specifications
Average Cycle Time |
28/38 minutes
Standard/Advanced cycles |
Temperature |
18°C-35°C |
Height
Width
Depth |
838.2mm
558.8 mm
812.8mm |
Weight |
125 kg |
Sterilization Chamber Configuration |
Rectangular |
Chamber Volume |
51.3 L * |
Dimension |
157 mm x 320 mm x 600mm (HxWxD) |
Two-Tiered Shelf |
313 mm x 597 mm (WxD) |
Door Width (Radius)/Angle |
558.8 mm/100° |
Cart Dimension (H/D/W) |
560 / 735 / 610 mm |
Cart weight |
38 kg |
Placement |
Countertop or cart-mounted for mobility |
Connections |
180-264 VAC, 47-63 Hz, 10 A |
Service Requirements |
In operation, the STERRAD® NX™ System should not be placed closer than 25 mm to a wall at the rear. The sterilizer should be installed in a space of sufficient size to permit access to all 4 sides of the system. Service access requires a minimum clearance of 61 cm above the top and 61 cm to the right of the system. The power receptacle should be positioned 30.5 cm to 61 cm above the floor. |
Operational Environment Temperature |
18°C to 35°C |
Humidity |
10% to 85% up to 30°C, linearly decreasing from 85% at 30°C to 70% at 40°C |
Height
Width
Depth |
560 mm
610 mm
735 mm |
Weight |
38 kg |
Independent Monitoring System (IMS) |
Enables remote, independent monitoring of all processes and is available as an option |
Barcode Reader |
Available as an option |
*Please contact your ASP representative to determine maximum capacity based on your facility's particular sterilization load configuration(s).