Total cycle time: Programmable to best match customers product and load configuration
Cycle temperature: 40°C to 65°C (104°F to 149°F)

External dimensions:
1120 mm (44.1 in) wide
1750 mm (69 in) high
1130 mm (44.4 in) deep

Weight:
Sterilizer-524 kg (1153 lbs)

Sterilization chamber:
0.21 m3 (7.2 ft3) usable volume

Shelf dimensions:
Three shelves: 595 mm x 595 mm x 101 mm
Shelf Strength: 25 kg (55 lb/shelf)

*Regulatory approvals:
- CAN/CSA-C22.2 No. 1010.1-92; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use
- UL 3101-1; Standard for Safety for Electrical Equipment for Laboratory Use
- EN 61010; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use.
- EN 55011; Group II Class A limits, based on CISPR 11; Emissions
- FCC 47 Part 18: FCC Requirements for Industrial, Scientific, and Medical Equipment
- IEC 601-1-2; Medical Electrical Equipment, Collateral Standard, Electromagnetic Compatibility
- CE, a mark of conformity for the European Community

Installation and Operating Requirements*

Connections:
208V, 3 phase, 50/60 Hz, 15 Amps, NEMA L21-20 twist lock outlet. Must have a dedicated circuit and separate neutral and ground conductors. The sterilizer requires a CBA phase rotation.

Cord length:
3m (10 ft)

Placement
Built-in wheels assure maximum mobility

Operational environment:
Temperature - 18°C to 30°C (64°F to 86°F)
Humidity - 10% to 85% relative humidity (noncondensing)

Installation space:
Minimum area - 1127 mm x 1130 mm (44.4 in x 44.5 in)
Surface - Flat = 5 mm/M (3/16 in/3 ft)
Minimum space for loading and unloading - 1300 mm (44.5 in)
Minimum service access - 1 M (3 ft) on all sides (unit can be moved for service)

Planned maintenance schedule:
6 months/1 year

 

 

 