STERRAD® 50 Specifications *
Total cycle time: 45 minutes
Cycle temperature: 40°C to 55°C (104°F to 131°F)
External dimensions:
644 mm (25 in) wide
1450 mm (58 in) high (with cart)
813 mm (32 in) deep
Weight:
Sterilizer-230 kg (500 lbs)
Cart-45kg (100 lbs)
Sterilization chamber:
Rectangular configuration 44 L (1.55 ft3) usable volume
Shelf dimensions-620 mm x 400 mm (24 in x 16 in)
Chamber height-178 mm (7 in)
Installation and Operating Requirements*
Connections:
120 VAC +/- 10% 20 Amps, dedicated NEMA L5-20 three wire grounding receptacle to supply
12 Amps continuous and 40 Amps momentary current.
Placement: Cart-mounted for mobility
Operational environment:
Temperature-15°C to 40°C (58°F to 104°F)
Humidity-10% to 80% relative humidity (noncondensing)
Installation space:
Minimum area-80 cm x 89 cm (31.5 in x 35 in); flat surface
Minimum distance from wall-80 mm (3 in)
Minimum service access-1 M (3 ft) on all sides (unit can be moved for
service)
Planned maintenance schedule: 6 months/1 year
*Regulatory approvals:
- CAN/CSA-C22.2 No. 1010.1-92; Safety Requirements for Electrical Equipment
for Measurement, Control, and Laboratory Use
- UL 3101-1; Standard for Safety for Electrical Equipment for Laboratory Use
- EN 61010; Safety Requirements for Electrical Equipment for Measurement, Control,
and Laboratory Use.
- EN 55011; Group II Class A limits, based on CISPR 11; Emissions
- FCC 47 Part 18: FCC Requirements for Industrial, Scientific, and Medical Equipment
- IEC 601-1-2; Medical Electrical Equipment, Collateral Standard, Electromagnetic
Compatibility
- CE, a mark of conformity for the European Community