200 System Specifications

Total Cycle Time 75 minutes

Cycle Temperature 115°F to 131°F (46°C to 55°C)

Dimensions and Weight

Width: 36 in (915 mm)
Depth: 44.5 in (1130 mm)
Height: 69 in (1754 mm)
Weight: 1047 lbs (475 kg)

Sterilization Chamber

Configuration: Rectangular
Usable volume: 5.3 ft3 (150 liters)
Two-tiered shelf: 23.5 in x 23.5 in (597 mm x 597 mm)

Installation and Operating Requirements

Placement:  In-wall or freestanding
In-wall options - single door or double door
Freestanding - single door
Built-in wheels provide mobility
Connections: 208 VAC/60 HZ/20A/3 phase/5 Wire WYE /NEMA L21-20 receptacle

Installation Space

Minimum Area: 36 in (W) x 44.5 (D) (91.5 cm x 113 cm)
Surface: Flat +/- 5 mm/m
Service Requirements:  1000 mm service side, 450 mm non-service side;
sides are selectable (39 x 18 inches);
single-door unit can be moved for service

Operational Environment

Temperature: 18°C - 40°C (64°F - 104°F)
Humidity: 10% - 85% Up to 30°C ambient and linearly decreasing from 85% at 30°C to 70% at 40°C.

Carriage Dimension (Optional Accessory)

Width: 30 in (762 mm)
Depth: 49.5 in (1257 mm)
Height: 35 in (889 mm)

*Regulatory approvals:
- CAN/CSA-C22.2 No. 1010.1-92; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use
- UL 3101-1; Standard for Safety for Electrical Equipment for Laboratory Use
- EN 61010; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use.
- EN 55011; Group II Class A limits, based on CISPR 11; Emissions
- FCC 47 Part 18: FCC Requirements for Industrial, Scientific, and Medical Equipment
- IEC 601-1-2; Medical Electrical Equipment, Collateral Standard, Electromagnetic Compatibility
- CE, a mark of conformity for the European Community

STERRAD® 200 Sterilization System
Download the following:
STERRAD® 200 Datasheet
STERRAD® 200 Installation Requirements
STERRAD® 200 Instrument Processing Guidelines