100NX™ System Specifications
The STERRAD® 100NX™ System requires no costly plumbing, water source, filters, heat boosters, or venting. Just plug it in and it’s ready to use. (Available in single- or double-door options.)
Technical Specifications
Total cycle time:
STANDARD cycle, 47 minutes
FLEX cycle, 42 minutes
Cycle temperature: Less than 131°F (55°C)
Dimensions and Weight:
Height 70.5 in (179.1 cm)
Width 30.5 in (77.5 cm)
Depth 40 in (101.6 cm)
Weight 910 lb (413 kg) single-door
962 lb (436 kg) double-door
Installation and Operating Requirements
Connections:
Three-Phase Power: 200-240 VAC, 380-415 VAC
(Line to Line) 50-60 Hz
Installation/Service Requirements
Service access requires a minimum clearance of 24 in (61 cm) above the top and 8 in (20.32 cm) on each side of the system.
The system should not be placed closer than 2 in (5.08 cm) from the rear wall.
Sterilization chamber:
Configuration Rectangular
Usable Volume Single Shelf: 2.9 ft3 (82.3 liters)
Double Shelves: 3.3 ft3 (93.4 liters)
Dimensions (WxDxH) 20.1 in x 28.9 in x 16.1 in (51 cm x 73.5 cm x 41 cm)
Two-Tiered Shelf (WxD) 17.3 in x 25.2 in (43.9 cm x 64 cm)
Operational environment:
Temperature-64°F to 95°F (18°C–35°C)
Humidity-10% to 85% (noncondensing)
External Options Supports external printer and provides flexible printout options